TA Instruments – Thermal Conductivity Meters (FOX Heat Flow Meters)
Thermal conductivity is crucial in determining a material’s ability to transfer heat, which is essential for optimizing energy efficiency and predicting thermal performance across various industries. Thermal property measurements also provide valuable information about a material’s composition, purity, and structure. The steady-state method directly measures thermal conductivity by applying a temperature difference across the tested material and measuring the average heat flux through it. The FOX Series of heat flow meters is designed to measure insulating materials and conforms to international standards such as ASTM C518, ISO 8301, and DIN EN 12667. It features solid-state heating and cooling of plates, thin film heat flux transducers, and four optical encoders, ensuring accurate measurement of heat flow for a wide range of sample dimensions.
Available in :
The FOX Series of heat flow meters is specifically designed to measure insulating materials, adhering to international standards such as ASTM C518, ISO 8301, and DIN EN 12667. The system employs solid-state heating and cooling of plates and thin film heat flux transducers to precisely control the temperature gradient across the sample. With multiple configurations accommodating a wide range of sample dimensions, the FOX Series ensures accurate measurement of heat flow for samples with dimensions representative of real-world conditions. Additionally, the incorporation of four optical encoders provides the utmost accuracy in measuring sample thickness. An Automatic Sample Feeder (ASF) is also available to maximize sample throughput and lab productivity.
Key Features
- Compact size and cost-effective system for thermal conductivity measurements
- Solid state heating/cooling for precise temperature control
- Optical encoder for the most accurate digital measurement of sample thickness
- Proprietary thin film heat flux transducers for the most representative sample heat flow measurement
Technical Specifications
| FOX 50 | FOX 200 | FOX 200 HT | FOX 314 | FOX 600 | FOX 801 | |
| Maximum Sample Thickness | 25mm (1″) | 51mm (2″) | 51mm (2″) | 100 mm (4″) | 203mm (8″) | 304mm (12″) |
| Square Sample Width[1] | – | 204 x 204mm (8″ x 8″) | 204 x 204mm (8″ x 8″) | 300 mm (12″ x 12″) | 610 ร 610 mm (24″ ร 24″) | 762x762mm (30″ x 30″) |
| Temperature Range | -10ยฐ C โ 110ยฐC FOX 50 0ยฐ C – 190ยฐC FOX 50 VHS | -20 ยฐC to 75 ยฐC | 25 ยฐC to 250 ยฐC | -20 ยฐC to 75 ยฐC | -10 ยฐC to 65 ยฐC | -10 ยฐC to 65 ยฐC |
| Temperature Resolution | ยฑ0.01 ยฐC | ยฑ0.01 ยฐC | ยฑ0.01 ยฐC | ยฑ0.01 ยฐC | ยฑ0.01 ยฐC | ยฑ0.01 ยฐC |
| Accuracy | ยฑ 3% | ยฑ2% | ยฑ2% | ยฑ1% | ยฑ1% | ยฑ1% |
| Reproducibility | ยฑ2% | ยฑ0.5% | ยฑ0.5% | ยฑ0.5% | ยฑ0.5% | ยฑ0.5% |
| Thermal Conductivity Range[2] | 0.1 to 10 W/(mK) (0.633 to 60.3BTU in/hr ft 2หF) | 0.005 to 0.35 W/mK (0.035 to 2.4 BTU in/hr f2 ยฐF) | 0.005 to 0.35 W/mK (0.035 to 2.4 BTU in/hr f2 ยฐF) | 0.005 to 0.35 W/mK (0.035 to 2.4 BTU in/hr ft 2ยฐF) | 0.001W/mK to 0.35 W/mK (0.007 to 2.4 BTU in/hr ft 2ยฐF) | 0.001W/mK to 0.35 W/mK (0.007 to 2.4 BTU in/hr ft 2ยฐF) |
| Available Configurations | Automatic Sample Feeder Vacuum | With External Thermocouple Kit: up to 2.5 W/mโขK | With External Thermocouple Kit: up to 2.5 W/mโขK | Automatic Sample Feeder External Thermocouple Kit Linear Gradient Guard | 0.001W/mK to 0.35 W/m K (0.007 to 2.4 BTU in/hr f+2 oF) With External Thermocouple Kit: up to 2.5 W/mโขK | 0.001W/mK to 0.35 W/m K (0.007 to 2.4 BTU in/hr f+2 oF) With External Thermocouple Kit: up to 2.5 W/mโขK |
| [1] Guard materials may be used to test specimens that are less than the nominal width [2] External Thermocouple Kit extends conductivity range to 2.5 W/mโขK | ||||||
| DTC 300 | |
| Method | Guarded Heat Flow Meter |
| Standard Test Method | ASTM E1530 |
| Specimen Compatibility | solids, pastes, liquids, thin films |
| Specimen Thickness | Maximum 1โฒ |
| Specimen Diameter | 50 mm |
| Temperature Range | -20ยฐC to 300ยฐC |
| Thermal Conductivity Range | 0.1 to 40 W/mK |
| Thermal Resistance Range (for each module) | 0.0005 to 0.010 m2K/W 0.002 to 0.020 m2K/W 0.01 to 0.05 m2K/W |
| Accuracy | ยฑ3% |
| Reproducibility | ยฑ1 to ยฑ2% depending on thermal resistance |
| Dimensions | 25โณ(63.5cm) W x 28โณ (71.12 cm) D x 37โณ (93.98 cm) H |
Available Models
More Products
TA instruments provides innovative material characterization instruments that are widely used for research, analysis, and quality control in the evaluation…